Cold Bend Test Chamber
The apparatus will be able to bend the sample backward and forward in a direction perpendicular to the plane of the conductors, the two extreme positions making an angle of 90� on both sides of the vertical.
Rate of flexing will be 60 per minute. A flexing is a movement through 180�.
A weight having mass of 0.5 kg will be provided with suitable grips for hanging to the sample.
The apparatus will have a provision for fixing the sample of tinsel cord (Flat) in the apparatus.
A 6 Digit reset able type counter with battery back-up (rechargeable) for retaining the information in case of power failure up to 72 hours will be provided for counting each flexing cycle.
The standard motors used in the apparatus.
The apparatus will be able to operate on single phase 230 V, 50 Hz. AC supply.
Technical Specifications
SIZE: As per requirement.
TEMPERATURE RANGE: (-) 40 Deg. C.
TEMPERATURE UNIFORMITY: +/- 1 Deg. C.
TEMPERATURE ACCURACY: +/- 1 Deg. C
CONTROL SYSTEM : PID (WITH PROGRAMMER)
|
|